D. Saraswathi *, G. P. Gigi , V. Niraimathi , A. Jerad Suresh
- Department of Pharmaceutical Chemistry, Madras Medical College, Chennai-600 003, India
Abstract
A simple, specific and accurate reverse phase high performance liquid chromatographic method was developed for the estimation of acebrophylline in capsule dosage form. A phenomenex Gemini C18, 5?m column having 250�4.6mm i.d. in isocratic mode with mobile phase containing diammonium phosphate buffer (pH 4):methanol (60:40) was used. The flow rate was 1mL/minute and effluents were monitored at 273nm. The retention time was found to be 4.54 minutes. The linearity was in the range of 80-120 ?g/mL. The proposed method was validated and successfully applied for the estimation of acebrophylline in capsule dosage form.